Exploring the Feasibility of Using Dense-Phase Carbon Dioxide for Sterilization

Originally Published MDDI May 2001STERILIZATION Though still in the research stages, sterilizing with CO2 may soon be a viable option—both economically and logistically—for medical device manufacturers.

May 1, 2001

49 Min Read
Exploring the Feasibility of Using Dense-Phase Carbon Dioxide for Sterilization

Originally Published MDDI May 2001

STERILIZATION

Though still in the research stages, sterilizing with CO2 may soon be a viable option—both economically and logistically—for medical device manufacturers.

Michael A. Matthews, Langdon S. Warner, and Heinz Kaiser

A basic schematic of the proposed CO2 sterilization chamber.

The conventional sterilization processes commonly used by medical device manufacturers—EtO, steam, gas plasma, and gamma irradiation—are proven technologies. By gauging the requirements of their products and doing cost comparisons for various methods, manufacturers settle on a sterilization process of choice. But a new sterilization process for medical devices, dense-phase carbon dioxide (CO2), may change some manufacturers' opinions. Although still in the research phase, sterilizing medical devices using dense-phase CO2 may prove to be a viable alternative both for OEMs and for end-users at healthcare facilities.

This article explores the technical and economic feasibility of a possible dense-phase carbon dioxide (CO2) sterilization process, comparing it to currently available commercial processes. Dense-phase CO2 has been shown to deactivate some microorganisms. Although research is ongoing, CO2 may prove to have certain technological, environmental, and safety advantages compared with existing sterilization methods.

For the purposes of this article, the costs of sterilization using available equipment were determined from a survey of operations in a major research facility. The capital and operating costs of the hypothetical CO2 sterilizer were estimated based on similar equipment that is currently being developed for other applications.

CO2BACKGROUND

The Supercritical Fluid State. When gaseous or liquid CO2 is heated and compressed above the critical temperature (31°C) and pressure (73 atm), it becomes a dense, highly compressible fluid that demonstrates properties of both liquid and gas. Substances in the supercritical fluid (SCF) state normally have better solvating ability than do the same substances in the liquid state. The SCF has a viscosity similar to that of a gas, but a density closer to that of a liquid. The properties of an SCF can be changed by adjustments of pressure and temperature. Because of the low viscosity of SCFs (compared with familiar liquid solvents), an SCF can penetrate small orifices. Additionally, substances disperse throughout the SCF rapidly, due to high diffusion coefficients. 1

Dense-Phase Fluids. The liquid phase of CO2 requires temperatures to be below ambient, but pressure can typically be decreased to between 700 and 1000 psi. The lower pressure decreases costs, but some of the unique properties of the SCF state are sacrificed as a result of the lower pressure. Nevertheless, the liquid state of CO2 conserves some of the solvent properties of the supercritical state—namely low viscosities, high diffusion coefficients, and acceptable solvent power. Collectively, the term dense-phase fluid refers to operations in either the supercritical or liquid states.

Environmental Benefits. CO2 is of particular interest in dense-fluid technology because it is inexpensive, nonflammable, nontoxic, odorless, and nonpolluting. Manufacturers—or healthcare workers and patients at the end-user stage—would not be exposed to toxic residues using CO2-based sterilization. Some current sterilization and disinfection methods—EtO processes, for example—use toxic and flammable substances that have a strong odor.

CO2 is not a hazardous substance regulated by the Environmental Protection Agency, and its use in sterilization would not contribute to the greenhouse effect because CO2 would be captured as a by-product from other industrial activity.

COMMON STERILIZATION METHODS

Steam. Steam sterilization is widely used by medical device manufacturers and end-user facilities because it is effective, efficient, safe, and inexpensive. It cannot be used to treat heat-sensitive materials, however. The main effect of steam is to denature the proteins in the microorganisms and break down the microbial proteins, lipids, carbohydrates, and nucleic acids.

EtO. EtO is a colorless, flammable, and highly reactive compound. Due to its low boiling point of 10.4ºC (50.7ºF) at atmospheric pressure, EtO behaves as a gas at room temperature. EtO chemically reacts with amino acids, proteins, and DNA to prevent microbial reproduction. 2,3 This alkylating agent is widely used to sterilize heat-, radiation-, and moisture-sensitive materials such as those containing plastics and microelectronics. It should not be forgotten, however, that EtO is toxic and explosive. EtO sterilization facilities require extra ventilation and safety precautions.

Gas Plasma. Like EtO, hydrogen peroxide plasma is used for the sterilization of heat-sensitive devices that cannot be treated above 60ºC. Plasmas can be created at high or low temperatures in strong electromagnetic fields. The advantage of using plasmas is the absence of toxicity after the electromagnetic field is shut down. 4 In this method, sterilization is achieved by the use of hydrogen peroxide gas plasma for 45 to 75 minutes total cycle time, at low temperatures and in an environment of low moisture.5

Method

Sterilizer

Number

Loads February

Loads March

Steam

Amsco 3053

5

499

606

Ethylene Oxide

Amsco 3017

5

82

93

Hydrogen Peroxide

Sterrad 100

1

15

73

Table I: Loads processed in the study facility's sterile processing department during February and March.

SURVEY OF STERILIZATION OPERATIONS

To provide a quantitative basis for comparing CO2 sterilization to existing technology, the sterilization operations at a major South Carolina hospital facility were surveyed. At this location, there are 48 machines performing sterilization that are dispersed among the different departments in the facility. The predominant sterilization method is steam, selected because of its cost advantages and effectiveness. The sterile processing department, however, has EtO and hydrogen peroxide gas plasma sterilizers in addition to the steam systems. These latter two methods of sterilization are used with heat-sensitive instruments.

Due to the diversity of the machines, chamber sizes, cycle times, rates of usage, and methods of sterilization, a cost per cubic foot was determined in order to compare costs among the different sterilization methods. The cost per cubic foot was determined by dividing the total monthly cost by the number of cubic feet processed in a month. The total monthly cost consists of the mortgage payment on the capital invested in the machines (with a fixed 7.5% interest rate for 10 years), the monthly portion of the respective maintenance cost, and the operational cost of processing each load in a given month.

The sterile processing department is the only department in the facility that sterilizes items with methods other than heat and steam. As a result, this department was more closely studied. Information about the number of loads was collected for the months of February and March 1999. Table I shows the number of loads during these two months and identifies which sterilization methods were used.

The steam loads were processed in Amsco 3053 machines, each with a capacity of 30 cubic feet. The EtO loads were sterilized with Amsco 3017 sterilizers, each with a chamber size of 5 cubic feet. The hydrogen peroxide gas plasma loads were processed in a Sterrad 100 with a 3.5-cubic-foot chamber. The cost comparison for the different sterilization methods in the sterile processing department is shown in Figure 1.

Figure 1. Cost comparison for the study's sterilization methods.

Personnel. The personnel cost for the sterile processing department was surveyed in order to obtain the personnel cost per cubic foot of sterilizing surgical instruments. At the time of the survey, there were three shifts. The function of these employees is to sort, disassemble, clean, assemble, inspect, pack, and wrap the items to be sterilized. This activity generates an additional dollar amount per cubic foot required for each of the different sterilization methods.

The personnel cost for processing heat-resistant devices is approximately $7 per cubic foot in the sterile processing department. Heat-sensitive devices have a higher personnel cost per cubic foot, however, at $37.

THE PROTOTYPE CO2 STERILIZATION PROCESS

Based on several patents, 6–12 as well as the experiments performed by several researchers,13–20 a hypothetical set of process conditions was chosen for a prototype CO2 sterilizer. The primary consideration is pressure, which may be as high as 200 atm or more in industrial SCF processes. The temperature required for supercritical CO2 sterilization is close to room temperature, ranging from 35º to 45ºC. Agitation of the CO2 is an important factor that may reduce the time required for sterilization. The water content may also affect the sterilization cycle time. Higher water content improves the permeability of an organism, allowing the CO2 to penetrate the cell wall and kill it. Although the method of sterilization using supercritical CO2 has not to date been completely defined by research, the general procedures are similar among the different researchers. This prototype CO2 sterilization cycle uses the following steps:

  • Prepare and package the materials.

  • Place materials into the high-pressure chamber.

  • Pump CO 2 at the selected pressure and temperature into the chamber.

  • Cycle the chamber pressure several times during the sterilization cycle.

  • Stop agitation and depressurize the chamber.

  • Remove the sterile material.

CAPITAL AND OPERATING COSTS

To determine the capital and operating costs of a hypothetical dense-phase CO 2 sterilizer, several manufacturers of existing commercial equipment were contacted, along with several experts in the field. Commercial applications include dry cleaning and parts cleaning with CO2. These processes operate at conditions similar to those expected for the CO2 sterilization process. The annual maintenance cost, according to equipment manufacturers, ranged from 3 to 10% of the cost of the equipment.21

The operational cost per load is determined by considering technical aspects of a CO 2 machine and the supplies required to comply with terminal sterilization (sterilization of sealed containers) standards. The variables affecting the cost per load are size of the chamber, degree of CO2 recycling, cost of CO2, cost of power, cycle time, and cost of supplies.

The Chamber. The chamber size determines the amount of CO 2 needed to sterilize or clean the load. In this study, the chamber size is 3.5 cu ft. The amounts of CO2 required for a 3.5-cu-ft chamber in the liquid and supercritical stages are 180 and 205 lb, respectively. Recycling minimizes CO2 losses, but also increases the power consumption. The recycling percentage selected for this analysis is 95%.

Although the cost of CO2 is relatively low ($0.20 to $0.25 per lb), it is an important factor in the cost per load. The CO2 cost used in this analysis is $0.20 per pound. The hourly operating cost for this analysis is $1.50.

Supplies. There can be no accurate information given about the cost of supplies for CO2 sterilization because these supplies are specific for each sterilization method. For example, the pouches and wraps used in the hydrogen peroxide gas plasma machine are different from those used in EtO and steam sterilization.

Wraps and pouches allow the sterilant to come in contact with the items being sterilized. Other supplies that must withstand the sterilization process include biological indicators, chemical indicators, and tape. For the purpose of this article, it is assumed that the cost of these supplies is similar to that considered in the EtO and gas plasma hydrogen peroxide methods. The cost per load of these supplies is approximately $8.25.

The cycle time influences the number of cubic feet that a sterilizer is able to process. Shorter sterilization cycles improve the turnover rate of the materials, offering more flexibility and decreasing the investment in additional equipment. A cycle time of 2 hours was selected for the hypothetical CO2 sterilizer.

The cost per cubic foot for CO 2 sterilization was determined following the previous model developed for steam, EtO, and hydrogen peroxide. The capital, maintenance, and operational costs were added, and this sum was divided by the maximum number of cubic feet that the CO2 sterilizer would process per year. The principal variables influencing the cost per cubic foot of CO2 sterilization, and the cost per cubic foot, are shown in Table II. This table depicts the conditions and values of probable equipment.

Variable

Parameter

Cost

Capital cost

Chamber size: 3.5 cu ft

$170,000

Maintenance

5% of capital cost

$8300/year

Operational cost

CO2 cost

Recycling 95% of 181 lb $0.20 per lb

$1.80 per run

Power cost

Cycle time 2 hours at $1.50 per hour

$3 per run

Supplies

Pouches, wraps, BI, CI, tape

$8.25 per run

Total cost per run

$13

Cost per cubic foot

$5.9

Table II: Variables affecting the cost of CO2 sterilization (prototype).

STERILIZATION METHOD COMPARISON

The steam sterilizer chosen as a basis for this study is an Amsco 3053-S prevacuum currently used in the facility's sterile processing department. This sterilizer has a maximum capacity of 30 cu ft for the sterilization of porous-, heat-, and moisture-stable goods. A microcomputer system monitors and controls all the operations and functions during the sterilization process.

The EtO sterilizer used for comparison is an Amsco 3017 that uses 100% EtO. This sterilizer, with a capacity of 5 cu ft, is equipped with an aerator to reduce the EtO levels in the breathing zone. A single-dose gas cartridge provides the 100% EtO, which is enough gas sterilant to process one load. This system also has a microcomputer that monitors and controls operational variables

The Sterrad 100 is a hydrogen peroxide gas plasma sterilizer produced by Johnson & Johnson. This sterilizer is able to process loads up to 3.5 cu ft. The installation and operating requirements consist of a simple connection to electricity; no water, steam, or air supply is needed. All of the operations are controlled and monitored by a microcomputer, which generates system performance records for each cycle. A cassette with 10 individual doses provides the hydrogen peroxide.

In the case of a prototype CO 2 sterilizer, it is assumed that a 3.5-cu-ft vessel is used. The vessel is constructed of stainless steel in order to support the elevated pressures needed to reach the supercritical stage (200 atm). The method of agitation is important, but the precise details must be determined through further research. CO2 machines must be designed with microcomputers that control and monitor the pressure, temperature, and time. For quality assurance and safety reasons, they must also register and print sterilization time, fluid conditions, and other parameters as is done with steam, EtO, and hydrogen peroxide sterilizers.

Hydrogen peroxide and CO2 sterilizers operate without process steam, air, or water. Moreover, they do not require drain or gas exhaust to dispose of their process waste. The two-hour CO2 cycle time is likely to compete with the Sterrad cycle of 75 minutes and is much shorter than the 15-hour cycle required by the EtO sterilizer.

COST BENEFITS OF CO2

In order to introduce a new sterilization technology, of course, it must be cost competitive. The technological advantages alone may not be enough to justify replacing capital equipment if the costs do not benefit customers and manufacturers alike. The capital costs of new technology are normally higher than those of established technology due to the investment made and the lack of economies of scale. Presumably, the estimated cost of the machine would decrease over time as a result of competition among different manufacturers and additional technological improvements.

Table III summarizes the cost comparison based on the assumptions and data described above. The maintenance cost of CO 2 sterilization is similar to that of a Sterrad 100; however, the cost of maintenance does not strongly influence the cost per cubic foot. The availability of service affects the decision of acquiring sterilizers. Service is an important factor for current customers of steam, EtO, and hydrogen peroxide gas plasma sterilizers. For example, both Steris and Johnson & Johnson offer a coast-to-coast service for repair or preventive maintenance. Other competitors offer similar service in the United States and abroad. Companies trying to commercialize CO2 sterilization would necessarily have to offer a similar service.

The cost per cubic foot for the different sterilization methods puts all these sterilization technologies on the same basis. Capital, maintenance, and operational cost are summed in a single cost. The cost per cubic foot for steam sterilization ($1 per cu ft) is the least expensive of all the methods considered in this analysis. This is as expected due to the advantages of steam sterilization previously mentioned. The cost per cubic foot of CO 2 sterilization is lower than that of EtO ($6 versus $19). Factors increasing the cost per cubic foot for EtO are the long cycle time, risk management activities, insurance, legal liability, and monitoring equipment, among others. Based on this cost analysis, CO2 and H2O2 sterilization would most closely compete due to the cost and technical advantages.

Cost

Steam

EtO

H2O2

CO2

Machine

70,000

40,000

123,000

170,000

Maintenance

1100

2500

8300

8300

Cost per load

25

77

17

13

Cost per cu ft

1

19

6

6

Table III: Cost comparison, in dollars, of the four sterilization methods.

CONCLUSION

CO2 sterilization may prove to be a viable substitute for current EtO sterilization of heat-sensitive materials and devices. This analysis of CO 2 sterilization shows a lower cost per cubic foot ($6) than EtO ($19) because of the shorter cycle time, lower cost per load, and lack of regulatory constraints. Moreover, CO2 does not have the potentially negative environmental and health effects of EtO.

Hydrogen peroxide gas plasma sterilizers have a shorter cycle time than prototype CO 2 sterilizers. Their cost per cubic foot is the same ($6), and both technologies have no impact on the environment or on manufacturing employees.

The present analysis is, of course, preliminary. Further research and development is required. CO 2 will have disadvantages. For example, the capital cost may be significantly higher than it is with existing technologies, even though the cost per cubic foot is lower than with EtO and equal to H2O2. Another practical problem would be the space needed to store the CO2 cylinders. Even though most of the gas is recycled, CO2 sterilization requires an inventory of several hundred pounds of CO2 per cycle. This could be a prohibitive constraint when comparing sterilization technologies.

CO 2-based technology is technically feasible for sterilization within the parameters suggested by available research. Equipment manufacturers from other applications (e.g., dry cleaning and precision cleaning) should be able to design and build a prototype that would satisfy the temperature, pressure, humidity, and agitation requirements for CO2 sterilization. CO2 is widely available and relatively inexpensive. Several companies distribute and sell CO2 in different grades.

The CO 2-based technology may be feasible and profitable if the application (market) is carefully targeted. The most likely market for CO2 among existing applications is as a replacement for EtO sterilizers. As noted, CO2 sterilization is less expensive than EtO sterilization. In addition, it does not pose the kind of environmental and health risks associated with the use of EtO. Hydrogen peroxide gas plasma sterilization is competitive with CO2 sterilization in that the two technologies have similar costs, their cycle times are approximately the same, and they do not produce toxic wastes.

Steam sterilization is the most effective method for the sterilization of heat-resistant instruments, and the costs of autoclaving are lower than those of CO 2 sterilization. It is not proposed that CO2 competes with steam for the sterilization of heat- and moisture-stable goods. But CO2 sterilization is a suitable method for the sterilization of heat-sensitive devices. The current trend in hospitals is to increase the number of less-invasive medical procedures, which are performed with complex devices that require low-temperature sterilization methods.

Finally, though the use of CO 2 sterilization offers cost and environmental advantages, there is no guarantee that a CO2 sterilizer will receive FDA approval. A sterilizer must achieve the sterility assurance level under the worst conditions, and it must be capable of meeting the sterilization specifications every time the process is performed. The process must maintain the functionality of the product and its packaging. All of these areas require further research.

REFERENCES

1.AD Russell, WB Hugo, and GAJ Ayliffe, Principles and Practice of Disinfection, Preservation and Sterilization (Blackwell Science, Oxford, UK: 1999).

2.PJ Landrigan, TJ Meinhardt, and J Gordon, "Ethylene Oxide: An Overview of Toxicologic and Epidemiologic Research," JAMA, 255 (1986) 1575.

3.DJ Hurrel, "Recent Developments in Sterilization Technology," Medical Plastics and Biomaterials, 5 no. 5 (1998) 26–37.

4.PT Jacobs Advances in Low-temperature Gas Plasma Sterilization Technology. Johnson & Johnson (1999).

5.G Park and A Crull, Supercritical Fluids: Their Time Has Come. (Business Communications Co., Norwalk, CT: 1994).

6.A Dillow, F Dehghani, J Hrkach, N Foster, and R Langer, Supercritical Fluid Sterilization Method. (PCT/US99/14266, Dec. 29, 1999).

7.Y Osajima, M Shimoda, T Kawano, and K Okubo, "Method for Inactivating Enzymes, Microorganisms, and Spores in a Liquid Foodstuff," U.S. Patent 5,667,835, Sep. 16, 1997.

8.Y Osajima, M Shimoda, T Kawano, and K Okubo, "Method for Modifying the Quality of Liquid Foodstuff," U.S. Patent 5,520,943, May 28 1996.

9.Y Osajima, M Shimoda, T Kawano, and K Okubo, "System for Processing Liquid Foodstuff or Liquid Medicine with a Supercritical Fluid of Carbon Dioxide," U.S. Patent 5,869,123, Feb. 9, 1999.

10.Y Osajima, M Shimoda, T Kawano, and K Okubo, "System for Processing Liquid Foodstuff or Liquid Medicine with a Supercritical Fluid of Carbon Dioxide," U.S. Patent 5,704,276, Jan. 6, 1998.

11.P Drumheller, "Materials and Methods for the Immobilization of Bioactive Species onto Polymeric Substrates," U.S. Patent 5,914,182, Jun. 22, 1999.

12.D Jackson, S Barcotta, and M Lepp, "Process for Cleaning, Sterilizing, and Implanting Materials Using High Energy Dense Fluids," U.S. Patent 5,213,619, May 25, 1993.

13.A Dillow, F Dehghani, J Hrkach, N Foster, and R Langer. "Bacterial Inactivation by Using Near- and Supercritical Carbon Dioxide," Rp. Natl. Acad. Sci. USA 96 no. 18 (Aug. 31, 1999) 10344–10348.

14.A Enomoto, K Nakamura, K Nagai, T Hashinoto, and M Hakoda, "Inactivation of Food Microorganisms by High-pressure Carbon Dioxide Treatment with or without Explosive Decompression," Biosci. Biotech. 61 no. 7 (1997) 1133–1137.

15.H Ishikawa, M Shimoda, H Shiratsuchi, and Y Osajima, "Sterilization of Microorganisms by the Supercritical Carbon Dioxide Micro-Bubble Method," Biosci. Biotech. Biochem. 59 no. 10 (1995) 1949–1950.

16.M Kamihira, M Taniguchi, and T Kobayashi, "Sterilization of Microorganisms with Supercritical Carbon Dioxide," Agric. Biol. Chem. 51 no. 2 (1987) 407–412.

17.H Kumagai, C Hata, and K Nakamura, "COCLASS=s10>2 Sorption by Microbial Cells and Sterilization by High-pressure CO2." Biosci. Biotech. Biochem. 61 no. 6 (1997) 931–935.

18.K Nakamura, A Enomoto, H Fukushima, K Nagai, and M Hakoda, "Disruption of Microbial Cells by the Flash Discharge of High-pressure Carbon Dioxide," Biosci. Biotech. Biochem., 58 no. 7 (1994) 1297–1301.

19.Taniguchi, M., Suzuki, H., Sako, M., Kobayashi, T. "Sterilization of Plasma Powder by Treatment with Supercritical Carbon Dioxide," Agric. Biol. Chem., 51, (12), 3425–3426 (1987).

20.CI Wei, MO Balaban, SY Fernando, and AJ Peplow, "Bacterial Effect of High Pressure CO2 Treatment on Foods Spiked with Listeria or Salmonella," Journal of Food Protection 54 no. 3 (1991) 89–193.

21.H Kaiser, "Technical and Market Feasibility of Using Dense Phase Carbon Dioxide for Sterilization in Hospitals," Master's Thesis, University of South Carolina, December 2000.

Michael A. Matthews, PhD, is a professor of chemical engineering at the University of South Carolina (Columbia, SC). He conducts research on dense-phase carbon dioxide. Langdon Warner, PhD, is a research professor and a consultant on pollution prevention and waste minimization. Heinz Kaiser is an industrial engineer who obtained his master's degree in earth and environment resources management at the University of South Carolina.

Copyright ©2001 Medical Device & Diagnostic Industry

Sign up for the QMED & MD+DI Daily newsletter.

You May Also Like