MDDI Online is part of the Informa Markets Division of Informa PLC

This site is operated by a business or businesses owned by Informa PLC and all copyright resides with them. Informa PLC's registered office is 5 Howick Place, London SW1P 1WG. Registered in England and Wales. Number 8860726.

Zimmer Biomet's Brain Surgery Robot Has a Software Problem

FDA flagged the company's recall of the Rosa Brain 3.0 system due to a software problem that incorrectly positions the robotic arm. At least one patient injury has been linked to the issue, but no related deaths have been reported.

Zimmer Biomet has received five complaints, including one patient injury, related to a software problem that can drive its Rosa robotic arm to an incorrect position during brain surgery.

Zimmer Biomet recalled its Rosa Brain 3.0 system due to a software problem that can drive the robotic arm to an incorrect position during brain surgery. The company initiated the recall back in September but FDA just flagged the recall on Thursday.

So far there have been five complaints related to the problem, including one patient injury, but no deaths have been reported.

Rosa Brain is a robotic platform designed to assist neurosurgeons in positioning medical instruments or implants during brain surgery. The device is composed of a compact robotic arm and a touch screen mounted on a stand. Different types of instruments may be attached to the end of the robotic arm depending on the procedure.

The company told FDA it would provide a label containing workaround instructions that can be applied directly to the robot. A Zimmer Biomet engineer is also being sent to each customer site to install updated software to correct the issue. Customers should have been contacted by Oct. 31 with information about the planned update and the estimated timing.