Medtronic NIM Vital Nerve Monitoring System Correction Deemed Class I

The device was recalled after it was found that the system had the potential for false negative responses.

Katie Hobbins, Managing Editor

August 7, 2024

1 Min Read
Medtronic NIM Vital Nerve Monitoring System
Image courtesy of Medtronic

Medtronic’s NIM Vital Nerve Monitoring System recall has been identified as Class I after it was found that the system had the potential for false negative responses. The recall is a correction, not a product removal.

The nerve monitoring system is used to locate, monitor, and stimulate the nerves of the skull during surgery, as well as the nerves that connect the brain and spinal cord to muscles or sensory cells. The system issues an EMG tone when the probe is placed on a nerve to help prevent nerve damage during surgery.

NIM is being corrected due to reports of false negative responses during procedures where the device may fail to issue an electromyography tone when the probe is placed on a nerve. Use of an affected device may result in nerve damage, facial nerve damage, nerve weakness, and nerve paralysis, according to Medtronic. Currently, there have been 10 reported injuries and no deaths associated with the recall.

The company sent an Urgent Medical Device Correction Letter to customers in June noting that the product is not required to be returned as software version 1.5.4 has been deployed to fix the issue.

About the Author

Katie Hobbins

Managing Editor, MD+DI

Katie Hobbins is managing editor for MD+DI and joined the team in July 2022. She boasts multiple previous editorial roles in print and multimedia medical journalism, including dermatology, medical aesthetics, and pediatric medicine. She graduated from Cleveland State University in 2018 with a bachelor's degree in journalism and promotional communications. She enjoys yoga, hand embroidery, and anything DIY. You can reach her at [email protected].

Sign up for the QMED & MD+DI Daily newsletter.

You May Also Like